# Web UI – Kalibrering

> Visa och redigera varje moduls kalibreringstabell (gain/offset) i webbläsaren, spara till modulen eller återställ fabriksvärden.

_Källa: E-Sharp (esharp.se) · Accordion Web UI_


The Calibration page shows the calibration table of each module on your station. Each table holds a gain and an offset for each calibrated value. You can edit the values, save them to the module, or restore the factory values.

## Accessing

In the side menu select **Calibration** (always present; if no module has a calibration table, the page says so). Firmware 6.0 or later is needed.

## What the page shows

The list on the left shows every module that has a calibration table (slot and name); click one to show its table. Above the table you see the module's **Product**, **Revision** and **Serial number**. The table has one row per calibration **Key**, with its **Gain** and **Offset**. You cannot add or remove rows — the table always has the keys the module defines.

## Edit values

Click a **Gain** or **Offset** cell and type a new value (comma or point both work). Changed cells are highlighted and the page shows how many rows changed. A non-numeric value is marked and blocks saving. **Undo changes** reverts to the values the module holds.

## Save to the module

Click **Save to module…**, read the dialog (product, serial number, number of changed rows) and click **Save**. The calibration is stored on the module — in its IDPROM, or in its CSV calibration file when the table does not fit. Saved values take effect at once; note the old values before saving if you may need them again.

## Restore factory values

Click **Restore factory values…** and confirm to return the module to its default calibration (its stored table is overwritten and measurements change at once).

> Each module is calibrated individually in production. Only change or restore its calibration when you know you need to, or when E-Sharp asks you to. Check the serial number in the save dialog to make sure you are changing the right module, and copy the current values before you edit — you cannot get them back after you leave the page.

## See also

- [Web UI – översikt](/accordion/support/docs/webui-oversikt)
- [Web UI – Channel Management](/accordion/support/docs/webui-channel-management)