Overview

A calibration table holds the data that converts between raw hardware values and engineering units (volts, amps, °C). Without calibration, readings may carry offset and gain errors; with it, measurements are accurate.

What a calibration table shows

A linear calibration is defined by a gain and an offset per key (y = kx + m). A table-based calibration maps several raw points to calibrated values, for example:

Raw value Calibrated value Unit
0 0.000 V
1024 1.250 V
2048 2.500 V
4095 5.000 V

The details shown include the channel name, calibration type (linear, table or polynomial), the gain/offset or coefficients, the valid raw range and the engineering unit.

Verifying calibration

Apply a known reference (for example 2.500 V from a calibrator), read the channel and compare — the difference should be within specification. For critical measurements, verify at several points (0 %, 25 %, 50 %, 75 %, 100 % of scale) and compute the error at each: error = (measured − actual) / actual × 100 %.

Types of error

  • Offset error — all readings shifted by a constant (correct with the offset term).
  • Gain error — error proportional to the value (correct with the gain/slope).
  • Non-linearity — error varies across the range (correct with multi-point or polynomial calibration).
  • Hysteresis — reading depends on direction (average up/down readings).

Calibration interval

Recalibrate critical measurements every 6–12 months, general use annually, and always after repair or after exposure to out-of-range conditions. Calibration requires traceable reference equipment (precision voltage/current sources, a calibrated DMM, and a temperature chamber for temperature sensors). Back up calibration files before any change, and verify after loading.