The Calibration page shows the calibration table of each module on your station. Each table holds a gain and an offset for each calibrated value. You can edit the values, save them to the module, or restore the factory values.
Accessing
In the side menu select Calibration (always present; if no module has a calibration table, the page says so). Firmware 6.0 or later is needed.
What the page shows
The list on the left shows every module that has a calibration table (slot and name); click one to show its table. Above the table you see the module’s Product, Revision and Serial number. The table has one row per calibration Key, with its Gain and Offset. You cannot add or remove rows — the table always has the keys the module defines.
Edit values
Click a Gain or Offset cell and type a new value (comma or point both work). Changed cells are highlighted and the page shows how many rows changed. A non-numeric value is marked and blocks saving. Undo changes reverts to the values the module holds.
Save to the module
Click Save to module…, read the dialog (product, serial number, number of changed rows) and click Save. The calibration is stored on the module — in its IDPROM, or in its CSV calibration file when the table does not fit. Saved values take effect at once; note the old values before saving if you may need them again.
Restore factory values
Click Restore factory values… and confirm to return the module to its default calibration (its stored table is overwritten and measurements change at once).
Each module is calibrated individually in production. Only change or restore its calibration when you know you need to, or when E-Sharp asks you to. Check the serial number in the save dialog to make sure you are changing the right module, and copy the current values before you edit — you cannot get them back after you leave the page.