Accordion™ is a modular build system for a test- and measurement interface, developed by E# to maximize reuse and flexibility for virtually any measurement need.
Introduction to Accordion™
Every piece of electronics requires some kind of test interface in order to be tested and verified. A test interface typically needs to:
- Provide power
- Connect to one or several communication buses
- Generate appropriate stimuli (analog or digital signals)
- Measure analog or digital signals
In addition to the DUT interface, it might also provide test fixture control — from a simple jig to a complex automated machine.
The challenge is that the products a test interface connects to are vastly different, even within one company. There is no “one size fits all” solution, so test interfaces are often designed in-house — time-consuming work that locks up skilled R&D engineers for months or years, with versions that become increasingly hard to maintain. A test interface is not only hardware: someone must also write software and firmware drivers, and often a complete GUI for troubleshooting.
With Accordion™, everything is a module: DUT connections, communication buses, fixture control, measurement, stimuli, subsystem control — even the software. The strength of the concept lies in the build system and the subsystem that connects all parts together. Common requirements (I2C, UART, USB, JTAG; power and analog measurement) are met with standard modules, and because modularity is at the heart of Accordion™, it can be adapted to virtually any functionality.
Architectural overview
At its core there is a bus system with defined functionality and power that allows for vastly different applications.
Module types
There are currently four module types in the Accordion™ build system:
| Module Type | Features | Comment |
|---|---|---|
| SO-DIMM | 260-pin, 100 user signals, several form factors | Most E# measurement modules are of this type |
| M-SLOT | Robust, high current | Power-hungry modules such as the PoE module and PSU module |
| N-slot | 100 user signals, very high density | Expands the AGENT™ top with new functionality |
| EXPAND | 24 V, differential I2C over M8 industrial connector, up to 30 m cable, cascadable | Functions suited to be placed outside the main unit (e.g. relays or PSUs in test fixtures) |
- SO-DIMM — four form factors (Type A–D); two can be stacked.
- N-SLOT — the N-TOP slots (four) on AGENT™ are primarily used for connector adaptation for measurement modules.
- EXpand™ — a hardware concept allowing versatile expansion outside the AGENT™ box, using the rugged M8 cable standard. Example EXpand™ devices: DMM Relay (isolated 32-channel mux), PSU Module (2-ch 24 V/5 A), Fixture Comm, Active Load (4-ch 0–24 V/0–1 A), PCIe Carrier, SO-DIMM Carrier. The EXpand™ frontend is based on LTC4331 (differential I2C, up to 60 m cable); in AGENT™ the I2C-LR module connects up to four downstream devices.
AGENT Q2™
The mechanical outline of AGENT Q2 is based on an extruded aluminium enclosure (1455Q2201 from Hammond), coloured blue with custom sides and lid. It can be installed inside or outside a test fixture; E# offers fixation methods, desktop stands and 19” rack mounts.
Build system & top board
Under the top cover, the board has many board-to-board connectors for customized N-slot modules. The N1–N4 slots break out the signals from the SO-DIMM slots; N5 is not connected to a SO-DIMM slot and provides communication buses (I2C, SPI, UART, JTAG) or fixture functions (fan control, light indications). On the other side, the T1–T3 SO-DIMM modules provide measurement capability to the N-slots (T1→N1, T2→N2, T3→N3; T4 on the base board). Two additional slots (M1/M2) provide power-related functionality (programmable PSUs, active loads, PoE).
Base board
The base board hosts subsystem functionality — control module, power generation and cooling — and connects to the top board via two PCIe riser boards. The BASE has four SO-DIMM4 slots; A1 holds the Control Module with the Raspberry Pi 4 (CM32, “32” = number of unique addressable I2C channels). System-internal buses (SPI/UART/JTAG) are distributed in a daisy-chain from the CM32, through the base slots, up to the AGENT top. AGENT BASE also provides power supplies, clocks, fan control, overcurrent limiting and the on/off button.
Software architecture
The software is divided into two parts: one runs on the AGENT’s on-board computer, the other communicates with the AGENT through a proprietary TCP protocol. On the application layer, a .NET wrapper provides an API for use with NI TestStand or LabVIEW, and a Python wrapper wraps the same .NET API.
The ‘Channel’ concept
All functions in an Accordion system are abstracted as a channel — a digital or analog input or output, a temperature input, a register, or a communications channel. When the system boots, a discovery process finds all attached hardware modules, and each module reports the channels it supports. A channel’s value is always represented as a string: booleans as "true"/"false", floating-point numbers with a dot decimal separator ("1.234"). Values are always in SI units where applicable (volts, °C, etc.).
Channel types
| Channel Type | Direction | Options |
|---|---|---|
| Analog | IN / OUT | Gain, Offset, RSE/DIFF |
| Digital | OUT | PushType, IOVoltage |
| Digital | IN | PullType |
| Pseudo-Digital | IN / OUT | VLOW, VHIGH |
| Temperature | IN | SensorType, Cold-junction destination, parameter |
| Multiplexer | — | |
| Resistance | IN | ClampVoltage |
| Counter | IN | |
| Frequency | IN | |
| Actuator | OUT | ToggleEnabled, ToggleTimeMilliSeconds |
| Powersupply | OUT | CurrentLimit, OvercurrentResponse |
| Register | IN / OUT | |
| Ratiometric | IN / OUT | Gain, Offset |
| UART | IN / OUT | Baudrate, BusType, TerminationByte |
| SPI | IN / OUT | ClockSpeed, SpiMode, ChipEnableHigh |
| I2C | IN / OUT | ClockSpeed |
| NumericResult | IN |
Highlights:
- Analog — Gain/Offset convert the value as y = kx + m (k = Gain, m = Offset). For outputs the reverse calculation is applied (model an external amplifier or attenuator).
- Digital — configurable PushType (OpenDrain, PushPull) for outputs and PullType (None, Up, Down) for inputs, with an IO voltage for push-pull.
- Pseudo-Digital — a digital channel backed by an ADC/DAC, with VLOW/VHIGH thresholds.
- Temperature — returns °C; the Thermocouple module can be configured for many sensor types (TC types J/K/E/N/R/S/T/B, RTD PT-10…1000, thermistors). Thermocouples need a cold-junction reference channel.
- Power Supply — enabling/disabling turns the output on/off; the value sets the output voltage; CurrentLimit and OverCurrentResponse (Off or Limit/constant-current) govern protection.
- Actuator, Resistance, Counter, Frequency, Register, Ratiometric, Multiplexer, NumericResult — each covers a specific abstraction (see the channel concept above).
Software modularity
Just like the hardware, the software and firmware are modular. Each hardware module has an EEPROM storing its type; on boot the subsystem scans all locations, loads the associated drivers and exposes their channels. Accordion Extension Modules (AEM) are software-only modules a user can load/unload at any time — E# has implemented AEMs for PMBus automation, MODBUS/CAN/custom buses, SCPI instruments, LED analysers (e.g. FEASA) and temperature chambers.
APIs
AGENT™ manages the connection as a TCP server accepting many simultaneous clients, so interactive debugging works even while TestStand is attached. The C# API ships for both .NET 4.8 (TestStand/legacy) and .NET 8, with four main categories: value get/set (GetValues/SetValues, string-based), channel configuration (ConfigureChannels), module management (LoadModule), and bus transactions (PerformBusTransaction, with specific interfaces such as II2cTransaction). The Python API is auto-generated from the C# API for consistency.
Contact