Web UI – Audio analyser

Gör en Tone Audio Module (ESH10000560) till en tvåkanalig ljudgenerator och analysator i webbläsaren — nivå, frekvens, THD+N, SNR, spektrum och scope.

Originaldokumentationen är på engelska. Visa som Markdown (för AI/KB)

The Audio analyser page turns a Tone Audio Module (ESH10000560) into a two-channel audio generator and analyser in your browser. You set a test signal on each output, choose which input each channel measures, and read level, frequency, THD+N and SNR live, together with a spectrum, a waterfall and a scope view. The module does the signal processing itself at 96 kSps and 24 bit — only settings and results travel between the module and your browser.

Accessing

Open the station in a browser (firmware 6.5 or later), then Instrument Panel → Audio analyser. The page appears when the station has a Tone Audio Module; each module has its own panel titled with its slot and name. The Analyser running switch starts and stops the measurements.

Channel readings

Each channel panel (CH1 left, CH2 right) summarises its settings and shows: Vrms and Vpp (input level at the pin), Frequency (strongest tone), THD+N (%), SNR (dB) and Clip (red when the input reaches the top of the range — the other readings are unreliable while it is red). The Output switch turns the channel’s output jack on or off; the generator keeps running while the output is off. The readings are also station channels (e.g. 0.2.ESH10000560.MEAS1_THD), so you can add them to the Chart Viewer.

Generator and input settings

Generator (per channel): On, Waveform (Sine, Square, Triangle, White noise, Pink noise, Sweep), Frequency (Hz), Duty (% for square), Level (% of full output), and Input to output. At 100 % level a sine is about 1.4 Vpp (0.49 Vrms). Input: which input pair (IN1–IN4) the channel measures, and Gain in dB (−6 dB normal; gain acts on IN1 and IN4 only). Measurement filters: a first-order low-pass (off, 24/20/16/10/5/2/1 kHz) and Weighting (Flat or A-weighted, applied to the spectrum, Vrms, THD+N and SNR).

Analysis view

The toolbar: Live (freeze/continue), View (Spectrum or Waterfall), per-channel show/hide, Averaging (off, 2/4/8 frames), Slope, Peak, Frequency axis (log/linear), Level unit, Harmonics (marks f0 and H2–H5 with levels), Hold reference, Reset zoom, and CSV/PNG export. Peak draws the highest level at each frequency (last 3/5/10 frames, Max hold, or Sweep response). The Waterfall view shows the spectrum over time, newest at the top, colour by level.

Scope and output waveforms

Turn on Scope to see the measured input signal in time (with DC and AC-RMS readouts) next to the signal the outputs are set to produce. Scope trigger (e.g. CH1 rising) steadies a repeating signal.

Measurement range and limits

Measurements are valid 50 Hz–20 kHz. All inputs and outputs are line level: on IN1/IN4 keep the input at or below 0.7 Vrms (0.85 Vrms max) at −6 dB gain; on IN2/IN3 at or below 0.35 Vrms (0.50 Vrms max). A square wave reads a high THD+N (≈47 %) by nature, and its edges ring slightly (≈9 % overshoot) from the converter’s filter — a square wave above ≈91 % of full scale can turn Clip red for the same reason. Terminate unused inputs when measuring low levels.

See also

Boka en demo av Probe

Visa oss ditt kort — vi granskar det och går igenom fynden tillsammans.

Föredrar du att mejla? info@esharp.se